Certifications for Temperature Probe Field Gauge LC20
LC Gauge ETL Authorization to Mark
LC Gauge EC Declaration of Conformity
LC Gauge IEC / IECEE CB Test Certificate (61010-1)
Ralston LC20 ACMA Declaration of Conformity (Australia)
Ralston PED Declaration of Conformity
REACH Declaration of Conformity
RoHS Certificate of Compliance
Field Gauge LC20 Registration of Broadcasting and Communication Equipment (Korea)| Weight | 12.4 oz / 350 grams |
|---|---|
| Length | 1.75 in / 45 mm |
| Width | 3.4 in / 87 mm |
| Height | 14.1 in / 36 cm |
| Battery Count | 2 |
| Battery Life | 1000 Hours |
| Battery Type | AA Alkaline (IEC-LR6) |
| Bluetooth Radio Installed | Yes |
| Calibration Certificate | ISO 17025:2017 Accredited |
| Certifications | Certified to CSA Standard C22.2 No. 61010-1 Complies with Electromagnetic Compatibility Directive 2014/30/EU Complies with Low Voltage Directive 2014/35/EU Complies with Pressure Equipment Directive 2014/68/EU |
| Contains Wireless Module | Yes |
| Country of origin | USA |
| Data Logging | 870 Readings (onboard) |
| Electrical Rating | 50 mA @ 5 VDC |
| Harmonized Code | 9026.20.0000 |
| Ingress Protection | IP67 (1 meter water submersion for 30 minutes) |
| LCD Display | 2.4 in / 61 mm segmented LCD Display |
| Materials | ABS / Polycarbonate Blend Powder Coated Aluminum |
| Max Operating Temperature | 122 °F / 50 °C |
| Min Operating Temperature | -4 °F / -20 °C |
| Panel Mount | No |
| Primary Power | Battery Power |
| Protection Class | Pollution Degree 2 (UL / IEC 61010-1) |
| Radio Approvals | FCC ID 2AA9B04 IC ID 12208A-04 Japan ID 210-107153 Korea - Clause 3, Article 58-2 of Radio Waves Act |
| Radio Type | Bluetooth 5 LE |
| Temperature Compensation | None |
| USB Connection | USB (Micro B) |
| Wetted Materials | 316 Stainless Steel |
| Wireless | Bluetooth |
| Wireless Range | 330 ft / 100 m |
| Thermal Probe | 1/8 in / 3.18 mm diameter, 12 in / 30.5 cm long, bottom mount |
| Measurement Type | Temperature |
| Media Compatibility | Air Alcohol Antifreeze Ethylene Glycol Hydraulic Oil Inert Gas Light motor oil Mineral Oil Natural Gas Petroleum Based Oil Ralston Calibration Oil Skydrol Transmission fluid Water Windshield washer fluid |
| Accuracy - Temperature Measurement | +/- (0.15 +0.002*|t|)°C (IEC 60751 Class A) |
| Temperature Units | K R °C °F |
| Max Probe Temperature | 302 °F / 150 °C |
| Min Probe Temperature | -22 °F / -30 °C |
| Sensor Bottom Thermal Probe | 1/8 in / 3.18 mm diameter, 12 in / 30.5 cm long, bottom mount |
| Sensor Bottom Wetted Materials | 316 Stainless Steel |
| Sensor Bottom Measurement Type | Temperature |
| Sensor Bottom Media Compatibility | Air Alcohol Antifreeze Ethylene Glycol Hydraulic Oil Inert Gas Light motor oil Mineral Oil Natural Gas Petroleum Based Oil Ralston Calibration Oil Skydrol Transmission fluid Water Windshield washer fluid |
| Sensor Bottom Accuracy - Temperature Measurement | +/- (0.15 +0.002*|t|)°C (IEC 60751 Class A) |
| Sensor Bottom Country of origin | USA |
| Sensor Bottom Temperature Units | K R °C °F |
| Sensor Bottom Max Probe Temperature | 302 °F / 150 °C |
| Sensor Bottom Min Probe Temperature | -22 °F / -30 °C |
| Sensor A Thermal Probe | 1/8 in / 3.18 mm diameter, 12 in / 30.5 cm long, bottom mount |
| Sensor A Wetted Materials | 316 Stainless Steel |
| Sensor A Measurement Type | Temperature |
| Sensor A Media Compatibility | Air Alcohol Antifreeze Ethylene Glycol Hydraulic Oil Inert Gas Light motor oil Mineral Oil Natural Gas Petroleum Based Oil Ralston Calibration Oil Skydrol Transmission fluid Water Windshield washer fluid |
| Sensor A Accuracy - Temperature Measurement | +/- (0.15 +0.002*|t|)°C (IEC 60751 Class A) |
| Sensor A Country of origin | USA |
| Sensor A Temperature Units | K R °C °F |
| Sensor A Max Probe Temperature | 302 °F / 150 °C |
| Sensor A Min Probe Temperature | -22 °F / -30 °C |
We’ll help you find the right solution for your application, whether you’re working in the field or a lab.